Marginal and internal fit of Cerec 3 CAD/CAM all-ceramic crowns.
Nakamura T, Dei N, Kojima T, Wakabayashi K. - Int J Prosthodont. 2003 May-Jun;16(3):244-8.
|
PURPOSE:
The purpose of this study was to examine the effects of the occlusal convergence angle
of the abutment and the computer's luting space setting on the marginal and internal
fit of Cerec 3 computer-aided design/manufacturing (CAD/CAM) all-ceramic crowns.
MATERIALS AND METHODS:
Mandibular second premolar all-ceramic crowns were fabricated for nine different
conditions using Cerec 3: all combinations of abutments with three different total
occlusal convergence angles (4, 8, and 12 degrees) with three different luting space
settings (10, 30, and 50 microm). The completed crowns were seated on the abutments,
and the marginal gaps were measured. The internal gaps between the crowns and
abutments were also measured, using test-fit silicone paste.
RESULTS:
When the luting space was set to 10 microm, the marginal gaps of the crowns
were greater than when it was set to 30 or 50 microm. When the luting space
was set to 30 or 50 microm, the marginal gaps ranged from 53 to 67 microm
and were not affected by the occlusal convergence angle of the abutment.
The internal gaps were within a range of 116 to 162 microm and tended to
decrease as the occlusal convergence angle of the abutment decreased.
CONCLUSION:
When the luting space was set to 30 microm, crowns with a good fit
could be fabricated on the Cerec 3 system, regardless of the occlusal
convergence angle of the abutment.
|
|
Related Articles
|
Influence of different convergence angles and tooth preparation heights on the internal adaptation of Cerec crowns.
Abstract
|
Margin gap size of ceramic inlays using second-generation CAD/CAM equipment.
Abstract
|
Marginal and internal fit of all-ceramic CAD/CAM crown-copings on chamfer preparations.
Abstract
|
Marginal fit and short-term clinical performance of porcelain-veneered CICERO, CEREC, and Procera onlays.
Abstract
|
An innovative method for evaluation of the 3-D internal fit of CAD/CAM crowns fabricated after direct optical versus indirect laser scan digitizing.
Abstract
|
Clinical and SEM evaluation of all-ceramic chair-side CAD/CAM-generated partial crowns.
Abstract
|
Evaluation of the marginal fit of three margin designs of resin composite crowns using CAD/CAM.
Abstract
|
Clinical marginal gap of porcelain fused to electro-formed gold coping crowns.
Abstract
|
Preliminary clinical results of a prospective study of IPS e.max Press- and Cerec ProCAD- partial coverage crowns.
Abstract
|
The significance of marginal gap and overextension measurement in the evaluation of the fit of complete crowns.
Free Full Text
|